Join Us, In Saving Lives And Engineering The Future
We are a diverse team of talented people who are dedicated to building technology that radically improves human lives.
The CerePlex M records distortion-free spikes and field potentials from high impedance microelectrodes with the flexibility and configurability for a variety of preparations such as freely moving rodents.
The small form factor of the CerePlex M provides a lightweight platform for neural recording in all configurations.
The CerePlex M, with Blackrock’s customized Honey Badger ASIC Chip, provides 7x higher resolution* to record high-fidelity transmission of extracellular spikes and local field potentials.
*Compared to competitors’ 12 Effective Number of Bits 2.44 μV vs. Blackrock’s 15.5 Effective Number of Bits 0.353 μV.
The CerePlex M features flexible ribbon cables allowing you to bend connectors in multiple configurations. Not only does this provide scalability of channels up to 128 but gives researchers the ability to record with different spatial configurations.
With a flip of a switch, the CerePlex M allows researchers to test electrode impedance while in the brain to monitor electrode tip condition. This helps researchers troubleshoot the quality and integrity of the electrode.
Spec | |
---|---|
Number of Channels | 32,64,96,128 |
Dimensions | 1" x 3/4" x 1/16" Wings added for additional channels |
Resolution | 16-bit ADC, 250 nV/bit @ normal gain |
Input Connector | Omnetics |
Compatible Array | Omnetics Utah Array |
Sampling Frequency | 30 kSps |
Input Frequency Range | 0.3 Hz - 7.5 kHz |
Input Impedance | 1300 MΩ @ 10 Hz, 13 MΩ @ 1 kHz |
Auto Impedance Measurement Range | 15 kΩ - 1 MΩ |
Maximum Input Range | +/- 8.192 mV with respect to reference |
Output Connectors | Micro HDMI |
Input Connectors | 1x, 2x, 3x, or 4x 32 Ch Omnetics, 36 Pin Omentics, 36 Position Dual Row Female Nano-Miniature (.025"/.64mm) Connector |
Project | LB# | Rev. |
---|---|---|
CerePlex M IFU | LB-0398 | 7.00 |