Join Us, In Saving Lives And Engineering The Future
We are a diverse team of talented people who are dedicated to building technology that radically improves human lives.
The NeuroPlex E is the next-generation of our FDA-cleared for human use digital headstage. It provides a noise-free link for high-fidelity transmission and recording of extracellular spikes and local field potentials.
The NeuroPlex E is FDA-cleared (K202174) for human use along with the NeuroPort Array, the only FDA-cleared microelectrode array, capable of recording hundreds of neurons from the human cortex. The NeuroPlex E gives researchers the power and flexibility to implement real-time control strategies such as restoring function.
The NeuroPlex E’s miniature form factor delivers a smaller and lighter, all digital, noise-free headstage for recording and stimulating
The NeuroPlex E converts analog signals to digital close to the recording site using the Honey Badger ASIC chip. This provides extracellular spikes and local field potential data that’s 7x higher resolution*.
*Compared to competitors’ 12 Effective Number of Bits 2.44 μV vs. Blackrock’s 15.5 Effective Number of Bits 0.353 μV.
| Spec | |
|---|---|
| Number of Channels | 128 |
| Dimensions | L 24.9 mm x W 17.7 mm X 17.9 mm |
| Resolution | 16-bit, at 250 nV per bit resolution |
| Compatible Array | NeuroPort Array |
| Sampling Frequency | 30 kS/s |
| Input Impedance | 1300 MΩ @ 10 Hz, 13 MΩ @ 1 kHz |
| Auto Impedance Measurement Range | 15 kΩ - 1 MΩ |
| Wide Band Mode | 0.02 Hz (1st order) - 10 kHz (3rd order) |
| Normal Band Filters: | 0.3Hz (1st order) - 7.5kHz (3rd order) |
| Maximum Input Range | +/- 8.192 mV with respect to reference |
| Project | LB# | Rev. |
|---|---|---|
| NeuroPlex E IFU | LB-0996 | 4.00 |
| Project | LB# | Rev. |
|---|---|---|
|
Warning: Trying to access array offset on false in /home3/blackry1/public_html/wp-content/themes/edesign/single-products.php on line 234 |
Warning: Trying to access array offset on false in /home3/blackry1/public_html/wp-content/themes/edesign/single-products.php on line 237 |
Warning: Trying to access array offset on false in /home3/blackry1/public_html/wp-content/themes/edesign/single-products.php on line 240 |
Nathan Copeland
BCI Pioneer | University of Pittsburgh